Micro Gear Pumps for Medical & Diagnostic Equipment
Brushless DC Gear Pumps

Micro Gear Pumps for Medical & Diagnostic Equipment

2026-09-05

Medical and diagnostic instruments often require controlled fluid delivery in a very limited installation space.

Typical applications include:

  • Reagent delivery

  • Sample preparation

  • Buffer circulation

  • Washing and flushing

  • Diagnostic fluid handling

  • Sampling

  • Automated liquid handling

  • Compact cooling circuits

The pump should be selected according to the complete fluid circuit rather than catalog flow alone.

1. Start With the Actual Fluid Circuit

An instrument may contain:

  • Reservoir

  • Filter

  • Tubing

  • Valves

  • Reaction chamber

  • Sampling chamber

  • Waste line

Every component contributes resistance.

Therefore, the pump needs to be evaluated at the actual operating flow and differential pressure.

2. Current BYMP Range

Boyang's current micro magnetic gear pump range includes:

Series Displacement Options Maximum Listed Flow
BYMP020 0.2 / 0.39 / 0.6 / 0.7 ml/rev Up to 2.2 L/min
BYMP030 1.0 / 1.2 / 1.7 / 1.9 ml/rev Up to 6.0 L/min
BYMP060 2.4 / 3.5 ml/rev Up to 10.5 L/min
BYMP070 4 / 6 / 9 / 12 / 17 ml/rev Up to 55 L/min
BYMP100 26 ml/rev Up to 65 L/min

For medical and analytical instruments, the smaller displacement configurations are often the first models to evaluate.

Higher-flow configurations can be considered when the instrument requires flushing, circulation or rapid fluid transfer.

3. Flow and Speed

Theoretical flow is approximately:

Flow = Displacement × RPM

Actual flow depends on:

  • Differential pressure

  • Viscosity

  • Temperature

  • Pump speed

  • Internal leakage

  • Pump configuration

This means an engineer should specify the required flow at the actual working condition.

4. Low Pulsation and Repeatability

Positive-displacement gear pumps can provide stable fluid delivery when properly selected.

However, actual flow stability depends on:

  • Gear design

  • Pressure

  • Speed

  • Fluid viscosity

  • Pump clearance

  • Motor control

  • Tubing and valve behavior

The complete fluidic circuit should therefore be tested rather than evaluating the pump as an isolated component.

5. Material Selection

Medical and diagnostic applications may involve reagents, cleaning fluids, buffers and other chemicals.

Potential wetted materials include:

  • SS316

  • Hastelloy

  • Special engineering plastics

  • Filled PEEK

  • Ceramic

  • Zirconia ceramic

  • FKM

  • EPDM

  • PTFE

  • FVMQ

  • CR

Material selection should be based on the actual chemical compatibility requirement.

6. Noise

Laboratory and diagnostic equipment is often used close to operators.

Current compact BYMP configurations can provide low-noise operation, with specified BYMP020, BYMP030 and BYMP060 straight-gear configurations below approximately 55 dB at 2000 rpm.

System-level noise still depends on motor, mounting and machine structure.

7. Suction and Sampling

Sampling equipment may place the fluid reservoir below the pump.

In these applications, inlet vacuum performance becomes important.

Current BYMP configurations can provide inlet vacuum capability around -0.85 bar, depending on speed and configuration.

Suction performance should be tested with the actual tubing, filter and fluid.

8. Dry Running During Startup

Some automated instruments may experience short periods of air ingestion during startup or reservoir replacement.

Current BYMP configurations are designed with materials and internal structures suitable for demanding operating conditions, but dry-running duration should always be evaluated for the specific pump, fluid and duty cycle.

Do not assume that a general dry-running statement means unlimited dry operation under every condition.

9. OEM Motor and Controller Integration

For medical and diagnostic equipment, the pump cannot always be evaluated separately from its motor.

The engineer may need:

  • 12 VDC or 24 VDC

  • PWM speed control

  • Closed-loop speed control

  • Encoder feedback

  • CAN or RS485

  • Compact wiring

  • Low electromagnetic interference

  • Defined startup behavior

Boyang can evaluate the pump and BLDC motor as an integrated OEM solution.

10. Prototype Validation

A practical validation process is:

Select → Prototype → Test Actual Fluid → Test Pressure → Test Temperature → Check Noise → Check Materials → Confirm Duty Cycle

Water testing is useful for initial mechanical evaluation.

Final validation should use the actual reagent, cleaning solution or working fluid.

FAQ

Which BYMP series is suitable for diagnostic equipment?

Low-flow applications can start with BYMP020 or BYMP030. BYMP060 and larger configurations can be evaluated for higher-flow transfer, flushing or circulation.

Can the pump materials be customized?

Yes. Materials, gears, seals, ports and motor configurations can be evaluated according to the fluid and instrument requirements.

Can the pump operate continuously?

Selected configurations can provide service life exceeding 20,000 hours under appropriate clean-medium conditions. Actual service life depends on the complete application.

Can Boyang supply the motor together with the pump?

Yes. Pump and BLDC motor parameters can be evaluated together for OEM integration.